Elcometer 130 SSP Soluble Salt Profiler
- Protective Coatings


The Elcometer 130 Soluble Salt Profiler provides fast and accurate measurement of the level and density of soluble salts - over 4 times faster than other Bresle equivalent methods.
- Summary
-
Summary
-
Four Bresle equivalent readings in 2½ minutes
Each filter paper is the size of four Bresle patches - the Elcometer 130 Soluble Salt Profiler not only displays the individual reading but also provides four Bresle equivalent readings in just over two minutes.
Tested under laboratory conditions in accordance with ISO 8502-9, the Elcometer 130 SSP provides equivalent measurements to the Bresle Patch Method. Click here to see the full report.
Detailed Analysis
In addition to soluble salt levels (cleanliness) or conductivity, the Elcometer 130 SSP Soluble Salt Tester carries out a detailed analysis of the test area - providing an accurate salt density profile map, pinpointing areas of high contamination outside user defined limits.
Generate instant reports with ElcoMaster® Data Management Software
The Elcometer 130 SSP Soluble Salt Meter wirelessly transmits readings, statistics and batches via Bluetooth® or via USB straight into an inspection application or into ElcoMaster® software, Elcometer ’s Mobile App, for instant report generation at your desk, or using your mobile in the field.
The new Elcometer 130 Soluble Salt Meter allows you to complete a Bresle equivalent test in just over two minutes.
The multi-point conductivity sensors enable the Elcometer 130 SSP Soluble Salt Tester to accurately display salt concentration, showing exactly where the contamination lies and generating full colour salt density maps in 2D or 3D.
Key features of the Elcometer 130 Soluble Salt Profiler include:
- Range of measurement modes:
- Cleanliness: two measurement options are available when measuring soluble salts – Bresle Method Equivalency or Elcometer 130 Equivalency mode
- Conductivity: ideal for measuring ionic contaminants, sometimes referred to as total dissolved solids –TDS, within a solution
- Multi-point conductivity sensor allows the trend in conductivity to be displayed as a density map
- Automatic temperature compensation ensures accurate results
- Large colour LCD screen displays readings in µg/cm², ppm, µS/cm, mS/cm, % salinity or mg/m²
- Stores up to 3,500 sets of readings in 1,000 alpha numeric batches
- Non-oxidising gold plated contacts ensure lifetime performance
- Transfer data via USB or Bluetooth® to ElcoMaster® for instant report generation
-
- Key Features
-
Key Features
Elcometer 130 SSP Soluble Salt Profiler
Accurate in all environments
The Elcometer 130 SSP Soluble Salt Meter has automatic temperature compensation ensuring accuracy in all climatic conditions. Impure water can be offset for accurate and repeatable readings.
Designed to last
Robust, durable & water resistant, the Elcometer 130 SSP Soluble Salt Profiler is available with a 2 year manufacturer’s warranty; giving you peace of mind.
Calibration verification - peace of mind
A Calibration Verification Tile is available for verifying the accuracy of the gauge whilst out in the field and the verification date is recorded for use in reports.
Display Modes
Large Single Reading
The reading value is displayed for the whole sample.
Four Bresle Patch Equivalent Readings
The measurement area is split into four Bresle Patch size quadrants and four Bresle Patch or Elcometer 130 equivalent readings are displayed simultaneously.
Pass/Fail to User Defined Limits
Provides a visual indication of salt contamination. Areas, where soluble salts levels are above the limit set, are shaded red, areas below the limit are white. The measurement bar to the left of the pass/fail map displays the limit (above the bar) and the percentage test area above the defined limit (below the bar). The bar is also a visual indication of the pass/fail percentage. The highest (Hi), lowest (Lo) and average (X) reading for the current measurement are displayed to the right.
2D Salt Density Map with High/Low Readings
Levels of soluble salts are illustrated using a green and red contour system; green for very low salt concentration or uncontaminated areas; yellow for areas with mid salt concentration; and red for areas with levels of salt contamination exceeding the defined map scale range.
3D Salt Density Profile & Peak Salt Concentration (Hi)
Provides a visual indication of the density of salt contamination over the test area.
How to use the Elcometer 130 Soluble Salt Profiler
Wearing clean disposable gloves (supplied), fill a syringe with precisely 1.6ml of pure distilled water.
Using tweezers, remove a filter paper from the pack and place it on the cleaned, non-labelled side of the magnetic disc supplied.
Disperse the water from the syringe, evenly across the whole of the filter paper and remove any bubbles from under the paper.
Place the magnetic disc, wetted face down, onto the area under test, pressing firmly into any contours or irregularities and start the 2 minute timer on the gauge.
After two minutes, carefully remove the filter paper and magnetic disc from the test surface and place them onto the measurement electrodes.
As each filter paper remains on the surface for two minutes, multiple tests can be undertaken at the same time, reducing inspection times further.
Close the lid, ensuring that the magnetic catch is fully engaged, the gauge will begin measuring.
The reading will be displayed on the screen in the chosen display mode.
If required for further analysis, place the filter paper in a resealable bag (supplied).
Elcometer 130 SSP Soluble Salt Meter: Made for iPhone 6 Plus, iPhone 6, iPhone 5s, iPhone 5c, iPhone 5, iPhone 4s, iPad Air 2, iPad mini 3, iPad Air, iPad mini 2, iPad (3rd and 4th generation), iPad mini, iPad 2, and iPod touch (5th generation).
“Made for iPod,” “Made for iPhone,” and “Made for iPad” mean that an electronic accessory has been designed to connect specifically to iPod, iPhone, or iPad, respectively, and has been certified by the developer to meet Apple performance standards. Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards. Please note that the use of this accessory with iPod, iPhone, or iPad may affect wireless performance.
iPad, iPhone, and iPod touch are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a trademark of Apple Inc., registered in the U.S. and other countries. Suitable for mobile devices running Android™ software version 2.1 and upwards. Android™ and Google Play are trademarks of Google Inc.
-
- Product Features
-
Product FeaturesElcometer 130 SSP Soluble Salt Profiler
Model SSP Repeatable & reproducible measurements Easy to use menu structure; in 30+ languages Tough, impact, waterproof & dust resistant; equivalent to IP64 Bright colour screen; with permanent backlight Scratch and solvent resistant display; 2.4" (6cm) TFT 2 year gauge warranty1 USB power supply; via PC Calibration certificate Calibration Verification Mode (with optional certified calibration tile) Ambient light sensor; with adjustable auto brightness Emergency Light Mode Magnetic & tripod mounting points Data output, USB to PC & Bluetooth® to PC, Android™ & iOS2devices On screen statistics Number of readings (n); Mean/Average, (x); Standard deviation (σ); Highest reading/Peak salt concentration (Hi;) Lowest reading (Lo); Coefficient of variation (CV%); Number of readings above high limit Gauge memory Number of individual reading sets; including salt density, pass/fail map & distribution graph 3,500Number of batches 1,000Measurement units & range Surface Cleanliness - Elcometer 130 Mode 0-50µg/cm², 0-500mg/m²Surface Cleanliness - Bresle Equivalent Method Mode 0-15µg/cm², 0-150mg/m²Conductivity 0-6000μS/cm, 0-6mS/cm
0-3000ppm, 0-0.3% SalinityResolution Surface Cleanliness 0.1μg/cm², 1mg/m²Conductivity 1μS/cm, 0.001mS/cm
1ppm, 0.0001% SalinityGauge Accuracy ±1% of readingMeasurement Mode Surface Cleanliness Conductivity Calibration Offset Mode Automatic temperature compensation ElcoMaster® software & USB cable Individual reading and profile map stored with time and date Plastic transit case Alpha-numeric batch names; user definable on the gauge Fixed Batch Size Mode; with batch linking Delete last reading Limits; (gauge & batch specific) Review, copy, clear & delete batches & calibration settings Trend graph; last 20 readings Batch review graph Analogue bar graph 1 The Elcometer 130 SSP is supplied with a 1 year warranty against manufacturing defects. The warranty can be extended free of charge to 2 years within 60 days of purchase
2 Visit www.elcometer.com/sdk to find out how to integrate Elcometer’s MFi certified products to your App.
-
- Technical Information
-
Technical SpecificationElcometer 130 SSP Soluble Salt Profiler
Part Number Description Certificate E130-SP Elcometer 130 Soluble Salt Profiler1 E130-SPC Elcometer 130 Certified Soluble Salt Profiler1 Operating Range 5°C - 40°C (41°F - 104°F) Power Supply 4 x AA batteries or via USB (rechargeable batteries can also be used) Battery Life Alkaline: Approximately 30 hours
Lithium: Approximately 45 hoursSample Time Sample Time Sampling Size 110mm (4.3”) circle Dimensions 250 x 145 x 50mm (9.8 x 5.7 x 1.9") Weight 780g (1.72lb) Packing List Elcometer 130 Soluble Salt Profiler, 3 x magnetic discs, 100 x high purity test papers, 250ml (8.5fl oz) pure distilled water, 20 x PVC storage bags, disposable gloves, sensor wipes, 3 x 3.0ml (0.1fl oz) syringes, 2 x plastic tweezers, 4 x AA batteries, shoulder strap, plastic transit case, test certificate and operating instructions, USB cable, ElcoMaster® software 1 Applicable Patents: GB 2527766
● Calibration Certificate supplied as standard.
-
- Standards
-
StandardsElcometer 130 SSP Soluble Salt Profiler
NSI 009-32, SSPC Guide 15, Bresle Equivalent ISO 8502-9 Test Method
-
- Downloads
-
Downloads
- Part Numbers
-
Part NumbersElcometer 130 SSP Soluble Salt ProfilerNo data found!!
-
Elcometer 130 Soluble Salt Profiler
- Part Number : E130-SP
-
Elcometer 130 Certified Soluble Salt Profiler
- Part Number : E130-SPC
- Accessories
-
AccessoriesElcometer 130 SSP Soluble Salt Profiler
- Part NumberT13027115
- Part NumberT13025964
- Part NumberT13024091
- Part NumberT13024093
- Part NumberT99911344
- Part NumberT13027596
- Part NumberT13024094
- Part NumberT13024092
- Part NumberT13024098
- Part NumberT13024087
- Part NumberT99920130
Welcome to Elcometer.tv
Your ultimate training channel -
- Home
- Coatings Inspection
- Coatings Inspection Division
- Surface Cleanliness
- Elcometer 130 SSP Soluble Salt Profiler
Elcometer 130 SSP Soluble Salt Profiler
- Protective Coatings
The Elcometer 130 Soluble Salt Profiler provides fast and accurate measurement of the level and density of soluble salts - over 4 times faster than other Bresle equivalent methods.
Four Bresle equivalent readings in 2½ minutes
Each filter paper is the size of four Bresle patches - the Elcometer 130 Soluble Salt Profiler not only displays the individual reading but also provides four Bresle equivalent readings in just over two minutes.
Tested under laboratory conditions in accordance with ISO 8502-9, the Elcometer 130 SSP provides equivalent measurements to the Bresle Patch Method. Click here to see the full report.
Detailed Analysis
In addition to soluble salt levels (cleanliness) or conductivity, the Elcometer 130 SSP Soluble Salt Tester carries out a detailed analysis of the test area - providing an accurate salt density profile map, pinpointing areas of high contamination outside user defined limits.
Generate instant reports with ElcoMaster® Data Management Software
The Elcometer 130 SSP Soluble Salt Meter wirelessly transmits readings, statistics and batches via Bluetooth® or via USB straight into an inspection application or into ElcoMaster® software, Elcometer ’s Mobile App, for instant report generation at your desk, or using your mobile in the field.
Summary
Elcometer 130 SSP Soluble Salt Profiler
The new Elcometer 130 Soluble Salt Meter allows you to complete a Bresle equivalent test in just over two minutes.
The multi-point conductivity sensors enable the Elcometer 130 SSP Soluble Salt Tester to accurately display salt concentration, showing exactly where the contamination lies and generating full colour salt density maps in 2D or 3D.
Key features of the Elcometer 130 Soluble Salt Profiler include:
- Range of measurement modes:
- Cleanliness: two measurement options are available when measuring soluble salts – Bresle Method Equivalency or Elcometer 130 Equivalency mode
- Conductivity: ideal for measuring ionic contaminants, sometimes referred to as total dissolved solids –TDS, within a solution
- Multi-point conductivity sensor allows the trend in conductivity to be displayed as a density map
- Automatic temperature compensation ensures accurate results
- Large colour LCD screen displays readings in µg/cm², ppm, µS/cm, mS/cm, % salinity or mg/m²
- Stores up to 3,500 sets of readings in 1,000 alpha numeric batches
- Non-oxidising gold plated contacts ensure lifetime performance
- Transfer data via USB or Bluetooth® to ElcoMaster® for instant report generation
Key Features
Elcometer 130 SSP Soluble Salt Profiler
Accurate in all environments
The Elcometer 130 SSP Soluble Salt Meter has automatic temperature compensation ensuring accuracy in all climatic conditions. Impure water can be offset for accurate and repeatable readings.
Designed to last
Robust, durable & water resistant, the Elcometer 130 SSP Soluble Salt Profiler is available with a 2 year manufacturer’s warranty; giving you peace of mind.
Calibration verification - peace of mind
A Calibration Verification Tile is available for verifying the accuracy of the gauge whilst out in the field and the verification date is recorded for use in reports.
Display Modes
Large Single Reading
The reading value is displayed for the whole sample.
Four Bresle Patch Equivalent Readings
The measurement area is split into four Bresle Patch size quadrants and four Bresle Patch or Elcometer 130 equivalent readings are displayed simultaneously.
Pass/Fail to User Defined Limits
Provides a visual indication of salt contamination. Areas, where soluble salts levels are above the limit set, are shaded red, areas below the limit are white. The measurement bar to the left of the pass/fail map displays the limit (above the bar) and the percentage test area above the defined limit (below the bar). The bar is also a visual indication of the pass/fail percentage. The highest (Hi), lowest (Lo) and average (X) reading for the current measurement are displayed to the right.
2D Salt Density Map with High/Low Readings
Levels of soluble salts are illustrated using a green and red contour system; green for very low salt concentration or uncontaminated areas; yellow for areas with mid salt concentration; and red for areas with levels of salt contamination exceeding the defined map scale range.
3D Salt Density Profile & Peak Salt Concentration (Hi)
Provides a visual indication of the density of salt contamination over the test area.
How to use the Elcometer 130 Soluble Salt Profiler
Wearing clean disposable gloves (supplied), fill a syringe with precisely 1.6ml of pure distilled water.
Using tweezers, remove a filter paper from the pack and place it on the cleaned, non-labelled side of the magnetic disc supplied.
Disperse the water from the syringe, evenly across the whole of the filter paper and remove any bubbles from under the paper.
Place the magnetic disc, wetted face down, onto the area under test, pressing firmly into any contours or irregularities and start the 2 minute timer on the gauge.
After two minutes, carefully remove the filter paper and magnetic disc from the test surface and place them onto the measurement electrodes.
As each filter paper remains on the surface for two minutes, multiple tests can be undertaken at the same time, reducing inspection times further.
Close the lid, ensuring that the magnetic catch is fully engaged, the gauge will begin measuring.
The reading will be displayed on the screen in the chosen display mode.
If required for further analysis, place the filter paper in a resealable bag (supplied).
Elcometer 130 SSP Soluble Salt Meter: Made for iPhone 6 Plus, iPhone 6, iPhone 5s, iPhone 5c, iPhone 5, iPhone 4s, iPad Air 2, iPad mini 3, iPad Air, iPad mini 2, iPad (3rd and 4th generation), iPad mini, iPad 2, and iPod touch (5th generation).
“Made for iPod,” “Made for iPhone,” and “Made for iPad” mean that an electronic accessory has been designed to connect specifically to iPod, iPhone, or iPad, respectively, and has been certified by the developer to meet Apple performance standards. Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards. Please note that the use of this accessory with iPod, iPhone, or iPad may affect wireless performance.
iPad, iPhone, and iPod touch are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a trademark of Apple Inc., registered in the U.S. and other countries. Suitable for mobile devices running Android™ software version 2.1 and upwards. Android™ and Google Play are trademarks of Google Inc.
Product FeaturesElcometer 130 SSP Soluble Salt ProfilerModel SSP Repeatable & reproducible measurements Easy to use menu structure; in 30+ languages Tough, impact, waterproof & dust resistant; equivalent to IP64 Bright colour screen; with permanent backlight Scratch and solvent resistant display; 2.4" (6cm) TFT 2 year gauge warranty1 USB power supply; via PC Calibration certificate Calibration Verification Mode (with optional certified calibration tile) Ambient light sensor; with adjustable auto brightness Emergency Light Mode Magnetic & tripod mounting points Data output, USB to PC & Bluetooth® to PC, Android™ & iOS2devices On screen statistics Number of readings (n); Mean/Average, (x); Standard deviation (σ); Highest reading/Peak salt concentration (Hi;) Lowest reading (Lo); Coefficient of variation (CV%); Number of readings above high limit Gauge memory Number of individual reading sets; including salt density, pass/fail map & distribution graph 3,500Number of batches 1,000Measurement units & range Surface Cleanliness - Elcometer 130 Mode 0-50µg/cm², 0-500mg/m²Surface Cleanliness - Bresle Equivalent Method Mode 0-15µg/cm², 0-150mg/m²Conductivity 0-6000μS/cm, 0-6mS/cm
0-3000ppm, 0-0.3% SalinityResolution Surface Cleanliness 0.1μg/cm², 1mg/m²Conductivity 1μS/cm, 0.001mS/cm
1ppm, 0.0001% SalinityGauge Accuracy ±1% of readingMeasurement Mode Surface Cleanliness Conductivity Calibration Offset Mode Automatic temperature compensation ElcoMaster® software & USB cable Individual reading and profile map stored with time and date Plastic transit case Alpha-numeric batch names; user definable on the gauge Fixed Batch Size Mode; with batch linking Delete last reading Limits; (gauge & batch specific) Review, copy, clear & delete batches & calibration settings Trend graph; last 20 readings Batch review graph Analogue bar graph 1 The Elcometer 130 SSP is supplied with a 1 year warranty against manufacturing defects. The warranty can be extended free of charge to 2 years within 60 days of purchase
2 Visit www.elcometer.com/sdk to find out how to integrate Elcometer’s MFi certified products to your App.
Technical SpecificationElcometer 130 SSP Soluble Salt ProfilerPart Number Description Certificate E130-SP Elcometer 130 Soluble Salt Profiler1 E130-SPC Elcometer 130 Certified Soluble Salt Profiler1 Operating Range 5°C - 40°C (41°F - 104°F) Power Supply 4 x AA batteries or via USB (rechargeable batteries can also be used) Battery Life Alkaline: Approximately 30 hours
Lithium: Approximately 45 hoursSample Time Sample Time Sampling Size 110mm (4.3”) circle Dimensions 250 x 145 x 50mm (9.8 x 5.7 x 1.9") Weight 780g (1.72lb) Packing List Elcometer 130 Soluble Salt Profiler, 3 x magnetic discs, 100 x high purity test papers, 250ml (8.5fl oz) pure distilled water, 20 x PVC storage bags, disposable gloves, sensor wipes, 3 x 3.0ml (0.1fl oz) syringes, 2 x plastic tweezers, 4 x AA batteries, shoulder strap, plastic transit case, test certificate and operating instructions, USB cable, ElcoMaster® software 1 Applicable Patents: GB 2527766
● Calibration Certificate supplied as standard.StandardsElcometer 130 SSP Soluble Salt ProfilerNSI 009-32, SSPC Guide 15, Bresle Equivalent ISO 8502-9 Test Method
Part NumbersElcometer 130 SSP Soluble Salt ProfilerNo data found!!-
Elcometer 130 Soluble Salt Profiler
- Part Number : E130-SP
-
Elcometer 130 Certified Soluble Salt Profiler
- Part Number : E130-SPC
Welcome to Elcometer.tv
Your ultimate training channelAccessoriesElcometer 130 SSP Soluble Salt Profiler- Part Number: T13027115
- Part Number: T13025964
- Part Number: T13024091
- Part Number: T13024093
- Part Number: T99911344
- Part Number: T13027596
- Part Number: T13024094
- Part Number: T13024092
- Part Number: T13024098
- Part Number: T13024087
- Part Number: T99920130
- Part Numbers
-